Errakal

Calibration regime fluid rig (small sizes)

The fluid rig has three primary standards:

1. Primary Standards

The principle of the two standards is different; The drum has a static start and stop.The Proverloop is a piston prover with a free running piston. The Proverloop has a "flying" start and stop. With a liquid turbine meter DN50 both standards are compared. Every five years (calamities or unexplained abnormalities earlier) is:

The weighing scale is calibrated every year.

2. Work Standards

In this regime is the following applied:

3. Regime for P and T measurements

For the relevant pressure and temperature sensors, a calibration takes place with an annual frequency.

4. Meters of third parties

The calibration of or research on flow meters of third parties are always against the reference standards. On request it is possible to calibrate the meters by 1 of the primary standards. The calibration of a flow point is measured three times. If there is a measurement out of line, to be judged by the operator, two additional measurements will follow. The judgement complies with the specifications concerning the reproducibility of the flow meter on the stand.

Key fields:

Specialties:

Site Map


ISO 9001
wibLinkedIn