Errakal

Gas flow calibration

Errakal has one rig, to calibrate all types of flow meters with air. See specifications below.

Description D ≤ DN 150 DN 150 < D ≤ DN 300
Flow 2,5 ÷ 7000 Nm3/h 2,5 ÷ 7000 Nm3/h
Medium air air
Gas temperature ca 20 ºC ca 20 ºC
Ambient temperature 20 ± 3 ºC 20 ± 3 ºC
Pressure Atm. ÷ 6,5 bar (a) Atm. ÷ ≤ 6,5 bar (a) *1
Available length 18 m 18 m
Working standards DN 50 CVM 1 x DN 150 turbine meter
Set of five turbine meters 2 x DN 80 turbine meter 2 x DN 100 turbine meter
Accuracy ≤ ± 0.25 % ≤ ± 0.25 %
Piping material Stainless steel Stainless steel
)1 Meters > DN 150 are not calibrated up to Q max at 6,5 bar (a). Q max can be reached at lower pressure levels, depending on Q max of the meter under test.

For the Reynolds area of the rig see graph.